System-sensor BEAM200 and BEAM200S Manuel d'utilisateur Page 6

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 15
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 5
STEP 1. COARSE ALIGNMENT
Refer to Figures 11 and 12 for this step.
1. Ensure that both of the optics lock-down screws are loosened so that the
optics will move freely.
2. Looking through the alignment mirror at both the alignment sight and
reflector simultaneously locate the position of the reflector in the optical
sight. This step will require some practice. It is necessary to train your
eyes to shift focus between the reflector and the mirror in order to locate
the reflector. If the distance between the reflector and the detector is
large it is helpful to place a brightly colored object on the wall near the
reflector to aide in seeing the reflector in the alignment mirror.
3. Once the reflector has been located, begin to adjust both the horizontal
and vertical alignment knobs so that the reflector becomes centered in
the alignment mirror. Take care in this step. If the optics are incorrectly
aligned in this step, it will not be possible to proceed with the fine adjust-
ment step.
STEP 2. FINE ADJUSTMENT
Refer to Figures 10 through 12 for this step.
In this step you will be fine-tuning the optics to the reflector. To provide feed-
back of the signal level coming from the reflector the dual digital display read-
out will be used. Due to the large distance range that the detector can operate
over it is necessary that the detector operate with many different settings of
electronic amplifier gain”. The detector is capable of determining the appro-
priate gain setting and then setting it itself via on-board processing algorithms.
There are no external gain settings on the detector that must be set by the
operator. Periodically throughout the fine adjustment step the detector will
need to re-adjust its “electronic amplifier gain” setting. When this occurs it
will be indicated by the dual digital readout as “- -”. When this occurs, cease
any further adjustment until the display again reads a number value.
1. Ensure that neither you nor any other objects are in the line of sight be-
tween the detector and the reflector.
2. Depress the Alignment switch once. Both the digital display and the yel-
low LED should turn on indicating that alignment mode has been en-
tered. The display should begin reading “- -” signifying an electronic gain
adjustment. After a few moments the display will indicate a numeric
value near 20. If the display reads “Lo” then the detector is not receiving
enough light from the reflector. Go back and repeat the course alignment
step and verify that the proper number of reflectors is used for the in-
stalled distance.
NOTE: The display will continue to read “Lo” until the detector receives
enough light from the reflector to continue with the fine adjustment step.
NOTE: In alignment mode (indicated by the yellow LED and the numeric
display) the sensitivity select and test switches are disabled.
3. With the display reading a numeric value, begin adjusting the horizontal
and vertical alignment knobs one at a time in the direction that increases
the numeric signal level on the display. Continue adjusting each axis
one at a time going back and forth between them until a peak value is
indicated. If a value of 90 is achieved, the detector will re-adjust the elec-
tronic gain once again. This will be indicated by a “- -” reading on the
display. When this happens halt any further adjustment until the display
again reads a numeric value. This process may occur more than once
during the fine adjustment step.
NOTE: Each time the display reads a value of 90 or greater the detector
will reduce the electronic gain. Each time the display reads a value of 10
or less the detector will increase the electronic gain.
4. Once satisfied that it is not possible to achieve a higher reading on the
display depress the alignment switch to complete the fine adjustment
step. The digital display readout will turn “OFF” and the yellow LED will
remain “ON”.
NOTE: It may not be possible to achieve a value near 90 on the display
during the last adjustment iteration. The final value of the display will
not likely be near 90. This is normal. It is due to the detector reducing
its electronic gain each time a value of 90 is achieved. When this occurs
the detector resumes with less electronic gain than previously when 90
was achieved. Less gain makes it more difficult to achieve higher values.
Final values anywhere between 20 and 90 are acceptable if no further
increase can be achieved.
NOTE: The alignment procedure is not complete yet.
At this time it is possible to set the sensitivity of the detector using the sen-
sitivity switch and digital display. See the Sensitivity Selection section of this
manual for further details.
STEP 3. FINAL GAIN ADJUSTMENT
Refer to Figure 13 for this step.
In this step, the detector will electronically adjust its internal gain one final
time. It is necessary to complete this step with the outer housing installed
since the housing will change the amount of light received from the reflector.
1. Tighten the optics lock down screws so the optics are secure.
2. Install the outer housing of the detector. The housing is installed by tight-
ening four screws, one in each corner of the housing. The screws are cap-
tivated in the plastic of the housing and cannot fall out during assembly.
NOTE: The housing contains a gasket seal that protects the detector
circuitry from corrosion and moisture sources. To Ensure that the gas-
ket seal performs correctly it is necessary to fully tighten all four of the
screws that hold the outer housing in place.
3. Remove the protective film from the front surface of the outer housing.
4. To initiate the final electronic gain adjustment, the reset switch must
be depressed. Once depressed the yellow LED will begin to blink. This
indicates that the detector is adjusting the electronic gain setting. Once
complete, the yellow LED will stop blinking and the green LED will begin
blinking. This indicates that the gain adjustment was successful.
NOTE: Use caution not to block the line of sight between the detector
and reflector in this step.
5. Install the outer aesthetic ring by snapping it onto the outer housing.
NOTE: If the outer aesthetic ring has been painted Ensure that the paint
is completely dry before proceeding with this step.
STEP 4. FINAL VERIFICATION
This step is required to Ensure the detector has been setup correctly and will
detect smoke at the proper sensitivity level.
1. With the detector functioning (indicated by the green LED blinking),
completely block the reflector with an opaque material. (Due to the high
optical efficiency of the reflector the selection of the opaque material
used to block the reflector is not critical. Acceptable materials include,
but aren’t limited to, this manual or the cardboard packaging inserts.)
See Figure 14. The detector should enter the fault condition (indicated by
the fault relay and the yellow LED (see Appendix I). If the detector does
not enter the trouble condition there is a problem with the installation
refer to the troubleshooting section in Appendix I for further assistance.
2. Complete a sensitivity test of the detector. Refer to the Sensitivity Testing
section of this manual for the appropriate procedure.
3. If the orange sticky paper was used to aid in the location of the reflector in
the alignment mirror it should be removed now. It is no longer necessary.
Congratulations. You have completed the final installation and alignment
procedure.
D400-74-00 6 I56-2289-005R
Vue de la page 5
1 2 3 4 5 6 7 8 9 10 11 ... 14 15

Commentaires sur ces manuels

Pas de commentaire